PRODUCT |  GENESYS 10S UV-Vis

Description

GENESYS™ 10S UV-Vis Spectrophotometer

Thermo Scientific™ GENESYS 10S UV-Vis spectrophotometer utilizes a high-intensity xenon lamp and dual-beam optical geometry to deliver unsurpassed data quality. Firing pulses of light only when the instrument is taking a measurement, the xenon lamp provides strong illumination from the UV to the near-IR region of the spectrum. Whether your application is research or routine analysis, educational or industrial, you can count on the GENESYS 10S for accurate and reliable results. The GENESYS 10S spectrophotometer series offer an excellent value providing high-performance, reliability, ease-of-use and a low cost of ownership.

 

Research Quality with Routine Simplicity :

- Accelerate through wavelength scans with scan speeds up to 3,600nm/minute

- Depend on dual-beam optics for superior photometric accuracy

- Acquire data from the UV to the near-IR

- Small footprint for easy transport and storage

- Increase sample throughput with the integrated 6-cell changer

- Thermostatting options with both circulating water and Peltier cooling

- Measure unusual or challenging samples with a variety of optional holders for test-tubes, long path cuvettes and filters

- Add a sipper accessory for easy sample handling


Feature-rich Local Control :

- Acquire, display, manipulate and save scanning, fixed wavelength, kinetics, or concentration data

- Save methods and data to an on-board library

- Start analyzing faster - save your most commonly used methods to the Smart Start menu and they - appear on screen when you turn the instrument on

- Archive methods or export data by saving to a USB memory device

- Perform automated Performance Verification with pre-programmed verification methods


Add Power and Capability with Thermo Scientific™ VISIONlite™ 5 Software :

- Stack spectra in Scan mode

- Measure at more wavelengths and do more calculations in Fixed mode

- Save and print your data as a formatted report

- Control the sipper accessory for automated use with volume-based pumping


Maintenance-Free Lamp :

- Save time with the instant-on xenon flash lamp

- Perform accurate analysis over the entire wavelength range of 190-1100nm

- Prevent damage to sensitive samples as the lamp only flashes when data is being acquired

- Save money with long-lifetime xenon flash lamp (guaranteed for 3 years)

- Lamp produces almost no heat so sample compartment temperature remains stable

 

Specifications :

Catalog Number

840-208100

840-208200

Beam Geometry

Dual Beam

Connections

USB Type A port for USB memory device (front panel)
USB Type B port for optional computer connectivity (rear panel)
USB Type A port for external printer (rear panel)

Container Type

Single cell holder and automated 6-cell changer standard

Data Interval for Scanning

0.1, 0.2, 0.5, 1.0, 2.0, 5.0 nm

Depth (English)

15.7 in.

Depth (Metric)

40cm

Description

GENESYS 10S UV-Vis

GENESYS 10S UV-Vis with printer

Detector Type

Dual Silicon Photodiodes

Display

Graphical with LCD backlight; 9.7 x 7.1 cm (3.8 x 2.8 in)

Drift

<0.0005A/hr

Electrical Requirements

Selected automatically, 100/240V 50/60Hz

Height (English)

9.8 in.

Height (Metric)

25cm

Hertz

50/60Hz

Includes

USB cable, USB memory device

Interface

Graphical with LCD backlight 9.7 x 7.1cm (3.8 x 2.8 in.)

Keypad

Sealed membrane with tactile response keys

Lamp

Xenon Flash Lamp (5 years, 3 years guaranteed)

Noise

<0.00025 at 0.0 A;
<0.00050 at 1.0 A;
<0.00080 at 2.0 A;
RMS at 260 nm

Optical Design

Dual Beam-Internal Reference Detector

Photometric Accuracy Instrument

±0.005A at 1.0A

Photometric Display

-0.1A - 4.0A; 11.5-125%T; ±99999C

Photometric Linearity

Up to 3.5Å at 260 nm

Printer

-

Internal printer:
40 column Internal (text and graphics)
Included with PNs 840-208100 and 840-209800
Optional accessory for PNs 840-208100 and 840-209700


External printer support:
External printer with USB connection and HP PCL 3.0 or higher printer language support

Scan Speed

up to 3600 nm/min

Spectral Bandwidth

1.8nm

Stray Light

<0.08%T at 220, 340nm (NaI, NaNO2) <1<1.0%T 198-200nm (KCl)

Voltage

100/240V

Wavelength Accuracy

±1.0 nm

Wavelength Range

190 to 1100nm

Wavelength Repeatability

±0.5 nm

Wavelength Scan Speed

4200nm/min

Weight (English)

19 lb.

Weight (Metric)

8.6kg

Width (English)

11.8 in.

Width (Metric)

30cm

 

 