GENESYS 50 UV-Vis Spectrophotometer, Cat. No. 840-298000
The GENESYS 50 is a dual beam UV-Vis spectrophotometer in a single-cell configuration that meets all of the specifications for critical quality control applications and precise chemical measurements.
Features:
- 7-inch high-resolution touchscreen user interface
- Local control for speed and reliability
- Networking capable (Wi-Fi ready)
- Large sample compartment with front access
Designed for :
- Chemistry labs
- Routine measurements
- Moderate throughput requirements
Specifications :
|
Product Size |
- |
|
Baseline Flatness |
±0.002A |
|
Compartments |
Accessible from top, front or side |
|
Connections |
Single USB-A supports flash memory devices for method and data storage Duplex USB-A on side supports connection to a Windows™ computer running optional remote control software, keyboard, mouse Export data to network or PC via USB, Ethernet, or Wi-Fi |
|
Description |
GENESYS 50 UV-Vis Spectrophotometer |
|
Detector Type |
Dual Silicon Photodiodes |
|
Dimensions (L x W) |
35.5 x 38.5 x 19.5 cm |
|
Display |
7-inch color touchscreen, fixed, high definition, 800 × 1280 pixels |
|
Drift |
(At 500 nm after 1 hour warmup) <0.0010A/Hr |
|
Electrical Requirements |
External AC to DC converter. Voltage and Frequency (Hz) selected automatically, 100–240 volts, 50–60 Hz. |
|
Keypad |
Touchscreen |
|
Lamp |
Xenon Flash Lamp (>5 years typical, 3 years guaranteed) |
|
Noise |
(RMS at 500 nm 60 consecutive measurements) ≤0.00020A at 0A at 260 and 500 nm ≤0.00030A at 1A at 260 and 500 nm ≤0.00040A at 2A at 260 and 500 nm |
|
Optical Design |
Dual beam |
|
Photometric Accuracy Instrument |
±0.002A at 0.5A ±0.004A at 1.0A ±0.008A at 2.0A |
|
Photometric Display |
-3A to +5A |
|
Photometric Range |
–2A to +3.5A |
|
Photometric Repeatability |
±0.001A at 1A measured at 1.0A at 546 nm |
|
Printer |
Snap-on printer available |
|
Resolution |
0.2 nm, 0.5 nm, 1 nm, 2 nm, 5 nm |
|
Spectral Bandwidth |
2 nm |
|
Stray Light |
< 1.0%T 198 nm (KCl) , <0.05%T at 220 nm (NaI), <0.03%T at 340 nm (NaNO2) |
|
Wavelength Accuracy |
±0.5 nm |
|
Wavelength Range |
190 nm – 1100 nm |
|
Wavelength Repeatability |
< ±0.2 nm |
|
Wavelength Scan Speed |
Slow, medium and fast - up to 1600 nm/min |
|
Weight (Metric) |
7.5 kg |
|
Warranty |